Straightness Measurement

Last Updated: 07/16/2012

Application Note Courtesy of Damalini

Straightness measurement is done on a large number of applications, for example balks, rolls, machine constructions, transportation lines etc.

The basic principle for straightness measurement

The basic principle for straightness measurement is that all measurement values will show the position of the detector unit relative the laser beam. To be able to use the values for adjustment and documentation it is needed to set absolute references (zero points). These can be points on the measured object or the horizontal level.

Horizontal level

When using horizontal reference, the laser beam is leveled according to the vials on the D22 laser transmitter. When the measured object is the reference, the laser beam is leveled to the detector positioned at the reference points. This leveling is always done according to the same principle: zero setting of the laser.

Straightness measurement with S and M units

It is also possible to perform a straightness measurement with the S and M units from a shaft alignment system (i.e. no separate laser transmitter is needed). The S unit is used as reference transmitter and the M unit is used as detector. In this case though, it is not possible to use horizontal reference.

Straightness measurement with first and last point as reference Straightness measurement with reference points moved
Straightness measurement with first and last measuring point as reference (EasyLink™). Straightness measurement with reference points moved (EasyLink™)

Filed under: Alignment